labtop-logo

Ice Lined Refrigerators

ice lined refrigerators and freezers

Labtop Ice lined Refrigerators are designed with temperatures range from 2.0Β°C to 8.0Β°C for safe and proper storage of vaccines as per WHO guidelines.. These refrigerators and freezers are made for extreme conditions and tested to perform in climates with ambient temperatures up to 43.0Β°C.

cabinet construction

The cabinet of Laptop Ice lined refrigerator is insulated with high density CFC free Poly Urethane Foam. Interior of chamber is Stainless steel (304, 0.8 mm thick). The exterior is white powder coated (1.0 mm thick) Mild Steel and the door has a magnetic gasket with a keyed lock. Stainless steel trays are provided inside.

safety features

  • Miniature circuit breaker to protect electronic components in case of abnormal power supply
  • Audio visual alarm when temperature deviates from set temperature
  • 180 seconds time delay for compressor switch on
  • Overload protector for safeguard the compressor

refrigeration system

The cooling is effected by CFC free refrigeration system with a condensing unit located inside the chamber. Labtop condensing units are characterised by high-efficiency heat transfer and low power consumption.
  • CFC free refrigeration with time delay to safe guard the compressor
  • Inner grooved copper tubes and slit aluminum fins for superior heat transfer
  • Filter drier for moisture and dirt protection
  • Safety and control devices for better system protection
  • Thermally protected and permanently lubricated fan motors

temperature control system

  • Micro-control based Temperature controller.
  • 4 digit, 7 segment LED display.
  • Audio visual alarm when the temperature deviates from the set temperature.
  • Password protected keypad lock.
  • Visual alarm for faulty Sensor.
  • High quality PT-100 Sensor for better accuracy.
  • Temperature alarm system with adjustable high-low alarm limits.
  • Display of Set Temperature and Process temperature.

hold over time

In the event of power failure, β€œholdover time” for any cold-chain equipment is defined as β€œthe time taken by the equipment to raise the inside cabinet temperature from its set temperature to the maximum temperature limit of its recommended range”, e.g. in the case of ILR, if the temperature is 2.0Β°C, then the time taken to reach 8.0Β°C from 2.0Β°C will be the holdover time for that ILR.

special features

  • Foamed in place CFC free PUF insulation ensures temperature stability and reduced power consumption.
  • Foamed in place CFC free PUF insulation ensures temperature stability and reduced power consumption.
  • Castor wheels for ease of mobility.
  • Adjustable leveling jack .
  • Meets WHO / UNICEF standards.
  • Thermometer to check the temperature on power failure.
Technical Specifications
Cat. No. LIR-75 LIR – 150 LIR – 300
Net Volume (liters ) 75 150 300
No. Of tray 2 2 5
Temperature Range (ΒΊC) 2.0 -8.0
Maximum ambient temperature (ΒΊC) 43.0
Accuracy(ΒΊC) Β± 0.5
Resolution (ΒΊC) 0.1
Hold over Time (hrs) 24 / 48 / 72
Refrigerant R-134a
Energy supply needed per 24 hours (hrs) 8
Voltage (VAC) 230 , 50/60 Hz
Power (W) 250

350

500
Product verification protocol E03/RF03-VP.1
Specification reference E03/RF03.1
Sensor Position 30 mm from the chamber
Operating conditions   4 – 45 ΒΊC and 95 % RH
Storage condition 0 – 50 ΒΊC and 95 % RH
Internal Dim (WxDxH)mm 450x420x420

600x430x400

960x500x500
External Dia (WxDxH)mm 550x520x825 1290x580x720 1650x782x755
External Dia with packing (mm) 680x650x1025 1420x710x925 1780 x900x960
Net weight (Kg) 150 250 300
Gross Weight (Kg) 175 300 350